| Technical Specifications | 310A | 410A | 510A | 610M | 120CW |
| Tissue/Vessel size | >60 μm | >60 μm | >60 μm | >60 μm / >450 μm up to 10 mm | >60 μm |
| Chamber(s) | Single bath | Dual, independent or common | Dual, independent or common | Four individual chambers | Single bath/conical shape |
| Chamber material | Acid-resistant stainless steel | Acid-resistant stainless steel | Acid-resistant stainless steel | Acid-resistant stainless steel | Acid-resistant stainless steel |
| Chamber volume | Max. 10 ml | Max. 10 ml | Max. 10 ml | Max. 8 ml | Max. 10 ml |
| Chamber suction | No | No | No | Manual or automatic, time controlled, user defined | No |
| Chamber cover | Yes | With connection for suction and gassing | With connection for suction and gassing | Supplied with connections for gassing | With connection for suction and gassing |
| Force range | -100 -/+200 mN | -100 - /+200 mN | -100 - /+200 mN | User selectable at +/-200/400/800/1600 mN | -100 - /+200 mN |
| Force resolution | 0.01mN | 0.01mN | 0.01mN | 0.01mN | 0.01mN |
| Micropositioners | Manually operated | Manually operated | Automatical linear motor operated | Manually operated | Manually operated precision micrometer |
| Weight calibration | Manual | Manual | Manual | Semi Automatic | Manual |
| Heating | Built into chamber, independent of superfusion | Built into chamber, independent of superfusion | Built into chamber, independent of superfusion | Built into chamber, independent of superfusion | Built into chamber, independent of superfusion |
| Temp. range | Ambient temp. to 50 °C | Ambient temp. to 50 °C | Ambient temp. to 50 °C | Ambient temp. to 50 °C | Ambient temp. to 50 °C |
| Temp. resolution | 0.1 °C | 0.1 °C | 0.1 °C | 0.1 °C | 0.1 °C |
| Temp. probe | External | External | External | External | External |
| Output reading | Force (mN) | Force (mN) | Force (mN) | Force (mN) | Force (mN) |
| Analog output | Up to four output, 1.0V full scale for all acquired signals, user defined | Up to four output, 1.0V full scale for all acquired signals, user defined | Up to four output, 1.0V full scale for all acquired signals, user defined | Independent filtered 4 channel output at 1.0V full scale | Up to four output, 1.0V full scale for all acquired signals, user defined |
| Serial output | Serial interface - RS232/RS485 | Serial interface - RS232/RS485 | Serial interface - RS232/RS485 | Serial interface - RS232/RS485 | Serial interface - RS232/RS485 |
| Voltage | 100 to 240 VAC (auto) 50/60Hz | 100 to 240 VAC (auto) 50/60Hz | 100 to 240 VAC (auto) 50/60Hz | 100 to 240 VAC (auto) 50/60Hz | 100 to 240 VAC (auto) 50/60Hz |
| Ambient temp. | 15-30 °C | 15-30 °C | 15-30 °C | 15-30 °C | 15-30 °C |
| Optional accessories | | | | | |
| - range | pH 0 - 14 | pH 0 - 14 | pH 0 - 14 | | pH 0 - 14 |
| - temp. correction | 0 - 50 °C | 0 - 50 °C | 0 - 50 °C | | 0 - 50 °C |