Technical Specifications | 310A | 410A | 510A | 610M | 120CW |
Tissue/Vessel size | >60 μm | >60 μm | >60 μm | >60 μm / >450 μm up to 10 mm | >60 μm |
Chamber(s) | Single bath | Dual, independent or common | Dual, independent or common | Four individual chambers | Single bath/conical shape |
Chamber material | Acid-resistant stainless steel | Acid-resistant stainless steel | Acid-resistant stainless steel | Acid-resistant stainless steel | Acid-resistant stainless steel |
Chamber volume | Max. 10 ml | Max. 10 ml | Max. 10 ml | Max. 8 ml | Max. 10 ml |
Chamber suction | No | No | No | Manual or automatic, time controlled, user defined | No |
Chamber cover | Yes | With connection for suction and gassing | With connection for suction and gassing | Supplied with connections for gassing | With connection for suction and gassing |
Force range | -100 -/+200 mN | -100 - /+200 mN | -100 - /+200 mN | User selectable at +/-200/400/800/1600 mN | -100 - /+200 mN |
Force resolution | 0.01mN | 0.01mN | 0.01mN | 0.01mN | 0.01mN |
Micropositioners | Manually operated | Manually operated | Automatical linear motor operated | Manually operated | Manually operated precision micrometer |
Weight calibration | Manual | Manual | Manual | Semi Automatic | Manual |
Heating | Built into chamber, independent of superfusion | Built into chamber, independent of superfusion | Built into chamber, independent of superfusion | Built into chamber, independent of superfusion | Built into chamber, independent of superfusion |
Temp. range | Ambient temp. to 50 °C | Ambient temp. to 50 °C | Ambient temp. to 50 °C | Ambient temp. to 50 °C | Ambient temp. to 50 °C |
Temp. resolution | 0.1 °C | 0.1 °C | 0.1 °C | 0.1 °C | 0.1 °C |
Temp. probe | External | External | External | External | External |
Output reading | Force (mN) | Force (mN) | Force (mN) | Force (mN) | Force (mN) |
Analog output | Up to four output, 1.0V full scale for all acquired signals, user defined | Up to four output, 1.0V full scale for all acquired signals, user defined | Up to four output, 1.0V full scale for all acquired signals, user defined | Independent filtered 4 channel output at 1.0V full scale | Up to four output, 1.0V full scale for all acquired signals, user defined |
Serial output | Serial interface - RS232/RS485 | Serial interface - RS232/RS485 | Serial interface - RS232/RS485 | Serial interface - RS232/RS485 | Serial interface - RS232/RS485 |
Voltage | 100 to 240 VAC (auto) 50/60Hz | 100 to 240 VAC (auto) 50/60Hz | 100 to 240 VAC (auto) 50/60Hz | 100 to 240 VAC (auto) 50/60Hz | 100 to 240 VAC (auto) 50/60Hz |
Ambient temp. | 15-30 °C | 15-30 °C | 15-30 °C | 15-30 °C | 15-30 °C |
Optional accessories | | | | | |
- range | pH 0 - 14 | pH 0 - 14 | pH 0 - 14 | | pH 0 - 14 |
- temp. correction | 0 - 50 °C | 0 - 50 °C | 0 - 50 °C | | 0 - 50 °C |